Title Of Article Chaper
Determination by gas chromatography of phosphine used in fumigation
Title Of Journal Book
Chemistry and industry
Pages
300-302
Collation
3 p. : ills., 2 figs.
Reference Bibliography
Includes bibliographical references
Language Of Text
English
Literature Type
Serial
Literature Level
Analytic
Abstract
A number of gas chromatograph detectors and columns were tested for determining concentrations of phosphine in air occurring during fumigation. For the range 0.02 to 3 mg/l the beta-ionization argon detector proved most suitable using the conditions described, giving results with an accuracy of ± 2 per cent. After standardization against dilute mixtures of phosphine in argon, concentrations can be read directly form a calibration chart to give results within minutes of sampling during fumigations. The thermionic (phosphorus) detector is more sensitive but requires very finely controlled condition more applicable to laboratory work.
pub_id
43068
Issue Date
19720000 Apr 1