Author
Fowles, M.; Wayne, R.P.
Title Of Article Chaper
Ozone monitor using an LED source
Title Of Journal Book
J. Phys. E.: Sci. Instrum
Volume
14
Issue
10
Pages
1143-1145
Literature Type
Serial
Literature Level
Analytic
Abstract
A description is given of a single beam photometer, for measuring ozone concentrations, that requires neither optical filters nor a mechnical chopper. Yellow-orange light-emitting diodes radiate a band whose maximum is shown to match closely the absorption maximum of ozone in the visible region, and the emission band is sufficiently narrow to ensure good Beer-Lambert behavior. The rapid rise and fall times of the diode permit electrical modulation of the light source. Over the emission band, the integrated absorption cross-section for ozone is about 4.2 x 10 <sup>25</sup>m<sup>2</sup>, while the peak ozone cross-section is only slightly greater (4.9 x 10 <sup>25</sup>m<sup>2</sup> at lambda