36211
Accession Number
27995
Title Of Article Chaper
Absorbed Dose from a Beta Source as Shown by Thermoluminescence Dosimetry
Title Of Journal Book
International Journal of Applied Radiation and Isotopes
Volume
28
Pages
625-627
Collation
3 p. : ill.
Reference Bibliography
Includes bibliographical references
Publisher
Pergamon Press
Language Of Text
English
Literature Type
Serial
Literature Level
Analytic
Abstract
A depth-dose curve was obtained for a <sup>90</sup>4SR- <sup>90</sup>Y beta source using a fine grain TL phosphor to measure the absorbed dose, aluminium absorbers being interposed between the source and the detector; the curve went through a maximum at an absorber thickness of about 40 mg cm<sup>-2</sup>. This curve was then used to predict the average dose rate to various thicknesses of calcium fluoride which has a similar absorption characteristic to aluminium; these values were compared with experimentally determined dose rates. This work was done in connection with thermoluminescence dating of flint and calcite in archaeology and geology.
Keywords
dose;absorption;beta;thermoluminescence;TL;dosimetry
pub_id
36211