Quantitative Analysis with Infrared Spectrophotometers: Differential Analysis
Title Of Journal Book
Analytical chemistry
Volume
24
Issue
4
Pages
619-622
Collation
5 figures
Reference Bibliography
Bibliography: p. 622
Literature Type
Serial
Literature Level
Analytic
Abstract
This work discusses the optimum conditions for quantitative analysis of two-component solutions if both components absorb at the wavelength at which the determination is made. Two conditions must be met for highest accuracy. (1) the solution as a whole should transmit about 40% of the incident energy. (2) When this condition is fulfilled, the interfering component should absorb as little as possible. In differential analysis with double-beam spectrophotometers the solution is placed in one beam and a known similar mixture is placed in the other. Increasing accuracy by changing the solutions in the beams is described. Constant slit widths are very important. This is more sensitive than single-beam methods.
Keywords
differential analysis;infrared spectrophotometer; quantitative analysis