10228
Accession Number
64
Title Of Journal Book
Evaluation of X-Ray Fluorescence Analysis for the Determination of Mercury in Coal
Collation
18 p.
Reference Bibliography
Includes bibliographical references
Report Number
NBSIR 75-675
Language Of Text
English
Literature Type
Monograph
Literature Level
Monographic
Abstract
Limits of detection for mercury in coal have been determined on both a wavelength-dispersive and an energy-dispersive x-ray spectrometer. They are between 2 and 3 ppm under best conditions for both spectrometers. Techniques for reducing the background intenisty measured by the energy-dispersive systems are discussed along with methods of preparing coal specimens for analysis in both instruments.
pub_id
10228