8203
Author
Olsson, Ingrid U.
Author Affiliation
Uppsala University. Department of Physics
Title Of Article Chaper
Routine checks in the Uppsala conventional <sup>14</sup>C laboratory to achieve reliable results
Title Of Journal Book
Radiocarbon
Volume
32
Issue
3
Pages
355-360
Collation
8 figs.
Reference Bibliography
refs.
ISSN
0033-8222
Language Of Text
English
Language Of Summary
English
Literature Type
Serial
Literature Level
Analytic
Abstract
Describes a series of routine self-checks that the Uppsala carbon-14 laboratory performs with all measurements; all uncertainties in the physical measurement of a sample are estimated. The laboratory studies long-term stability, calculates mean values for oxalic acid and background, and compares expected and real statistical distributions of uncertainties. To reduce the risk of bias, the samples from each series are almost always run on the same counter. Some samples are, however, run on two or more counters to check the possible bias to achieve reliable activity comparisons with other laboratories. It is always possible to trace which counter is used, since different number series are used for different counters.
Keywords
laboratories;carbon-14;stability;oxalic acid;quality control; radiocarbon dating;reliability;methodology;error